Electronic flowmeter
High precision flowmeter, instantly know the fresh gas flow to patient.
O2 and N2O linkage device ensure O2 concentration no less than 25%.
12.1″ TFT LCD touch screen
Displays the Ventilation parameters, Alarm information and Oscillogram.
High sensitivity touch screen ensures accurate and easy operation. Alternate button for dual control.
Technical Specification:
<td valign="top" width="243" style="box-sizing: border-box; margin: 0px; padding: 9px 12px; border-top: none rg
Ventilation modes: | P-CMV, V-CMV, P-SIMV, V-SIMV, PSV, SIGH, MANUAL |
Ventilator parameter ranges | |
Flow meter | O2(0.1-10L/min) |
N2O(0.1-10L/min) | |
AIR(0.1-10L/min) | |
Rapid oxygen supply | 35L/min-75L/min |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~ 30 cmH2O |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 60 cmH2O |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Oxygen concentration | 15 % ~ 100 %
|
Graphical display: | |
P-T(pressure – time) | |
F-T(flow – time) | |
P-V loop (pressure – volume loop) | |
V-T(volume – time ) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100% 18%-99% |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit
| Show on screen Show on screen |
Working conditions | |
Gas source | O2,N2O,AIR |